|

Measurement Accuracy Requires a Calibration
Program
When you use a measurement instrument, you
assume it is giving you an accurate measurement.
-
However, that is not always the case. Even
normal use and wear of the device can affect its
measurements.
-
Combine normal wear and tear with the
realities of a manufacturing environment (such
as oil, metal or plastic chips, and the chance
that the gage or instrument may be dropped or
jolted) and you start to realize the importance
of a good calibration system.
A formal measurement device management program
including a rigorous calibration program is
required to ensure measurement accuracy.
Components of a good measurement device
management program include:
-
Instrument inventory lists
-
Assessment of testing needs
-
Calibration procedures including the use of
contracted calibration services if needed
-
Records of device maintenance history,
calibration records, and traceability
documentation
-
Management of calibration standards
-
Identification of training requirements &
related resources
-
System audit requirements
Calibration Systems
-
ISO 10012-1 (titled, "Quality Assurance
Requirements for Measuring Equipment, Part 1,
Metrological Confirmation System for Measuring
Equipment) and ANSI/NCSL Z540-1 define the
requirements for the establishment and
maintenance of a calibration system designed to
control the accuracy of measuring and test
equipment and measurement standards.
-
A
Calibration Record should include:
-
Description of the device
-
Characterization and condition of the
measuring instrument
-
Date of calibration
-
Calibration procedure
-
Environmental conditions during calibration
-
Measurement and results supported by data
-
Identification of any failures
-
A traceability statement
-
A statement of the estimated uncertainty
-
It is important to have written procedures for
calibration procedures. Steps to check the
calibration of a device include:
-
Define the temperature and relative humidity
requirements.
-
Visually examine the device for damage and
wear.
-
Clean the measuring surfaces being careful not
to damage them.
-
Check the device using the appropriate
standards. If the device has a variable scale,
check the low, mid-point, and high-end of the
range.
If checks indicate the measuring device is not
accurate, remove the device from service.
-
Initiate action to adjust and recalibrate if
possible.
-
If the device cannot be fixed, tag it with an
"inactive" sticker and destroy it if that is
appropriate.
If checks indicate the measuring device is
accurate, attach the appropriate calibration
sticker and return the device to service.
 |
|